Glove Box With Gas Purification System

Application: MR-Lab2000 Single-sided Double-station Glove Box is a high-performance, high-quality closed cycle working system that automatically absorbs water and oxygen molecules to ultimately keep the environment inside the box clean. This system can provide an inert atmosphere environment with O2 and H2O within 1ppm. It is an economical circulation purification system designed to meet customer scientific research and development.
Send Inquiry

Product Introduction

Dual-Workstation Single-Side (2400*750mm) Vacuum Glove Box with Single Column Purification System

2 work station glove box

mR-Lab2000-2400*7

 

Application: MR-Lab2000 Single-sided Double-station Glove Box is a high-performance, high-quality closed cycle working system that automatically absorbs water and oxygen molecules to ultimately keep the environment inside the box clean. This system can provide an inert atmosphere environment with O2 and H2O within 1ppm. It is an economical circulation purification system designed to meet customer scientific research and development.
 

Features

 

* Closed cycle, the inert gas in the glove box circulates through the circulating fan and purifier to continuously remove water and oxygen;
* Automatically control regeneration, water and oxygen removal materials can be regenerated, and the regeneration process is controlled by the program;
* Automatic cleaning, the atmosphere replacement in the glove box is completed through the automatically controlled cleaning valve;
* Box pressure control, the pressure in the glove box is automatically controlled by PLC;
* Control, can be set freely, the system automatically protects when exceeding +/- 12mbar;
* The vacuum pump is automatically controlled and the vacuum pump can be automatically turned on when the system needs it.

 

Specifications

 

Structure

Integrated structure (Purification System integrated under the box)

Power Supply

110 - 220V AC, 50/60Hz, Single-phase

Glove box Box

Box Material: 304 Stainless Steel (Type 304), thickness 3mm

Box Size(excluded antechamber and support): 2400mm(L) x 750mm(W) x 900mm(H)

94.49 " x 29.5" x 35.4"

Box Support height: 900mm(35.43" )

Tray

Standard Layout

product-229-160

Two layers in box.

 

Option/Add: tool boxes or other

product-229-160

 

Various tool boxes, can be placed according to user needs.

Front window

Tilt design operating window, removable safety toughened glass front window, 8mm thickness,Wear resistance, corrosion resistance, good light transmission, sealing ring using 3/8 inch thick OMEGA sealing ring.

Option: Window Open

The front window can be designed to open instead of fix, in order to put in a variety of instruments (custom function).

product-229-211

Environment of Box inside

H2O <1 ppm (20°C/68 ℉, 1 atm)

O2 <1 ppm (20°C/68 ℉, 1 atm)

Antechamber

Control way: manual

Design: two circular antechambers

Main chamber size: Φ360mm (Dia.)x 600mm(length)(location: right-side of box)

14.17 "(Dia.) x 23.62" (L)

Mini chamber size: Φ150 mm(Dia.) x 300mm(L)(location: right-side of box)

5.9 "(Dia.) x11.8" (L)

Vacuum degree:<=-0.1MPa(right side antechambers included main chamber and mini chamber, all of that are need vacuum and display. The vacuum pressure gauge brand is German WIKA。

Option/add auto control way

product-229-160

It will automatically control the antechamber by time setting and the number of times for gas exhaust and filling.

Option/Add: heating mode

product-229-160

Max temperature:200℃/ 392℉

Control accuracy:±2℃/±3.6℉

The main chamber and mini chamber are both can adding the heating function (compared with standard antechamber the size is a little different)

Working gas and regeneration gas

Working gas: 99.999% high pure N2, Ar, He and other inert gas

Control gas:99.999% high pure inert gas

Regeneration gas: mixing gas of working gas and H2,hydrogen(H2)Volume accounted for 5-10%

If the pure system only has moisture removal function, the regeneration gas is the same with working gas

Gas purification system

Single Column Gas Purification System included:

German BASF Oxygen removal material, US UOP high moisture absorb material. The regeneration process of purification system is auto control and automatic moisture and oxygen removal function.

Gas purity can be sustained and keep long-term:H2O<1ppm, O2<1ppm。

Option: dual columns purification system

(one in use and the other is reserve)

Why choose duel pure column?

1、The Pure column need to be regenerated after saturated by absorbing the oxygen and moisture. The regeneration time is 16hours. If there is no reserved column, it may stop working. (we recommend dual column purification system to factory or other unit which need glove box working Continuously);

2、The pure capacity of dual pure columns is two times as much as single column

Pressure control system

Box pressure can be auto controlled by PLC touch screen

Control accuracy:±1Pa

Can also be manually controlled by foot switch

Filter system

The ends of inlet and outlet both are provided with filter

The filter precision is: 0.3 mu m

Closed circulating fan

High speed frequency conversion fan, flow 0-100m3/h, microprocessor control, Stainless steel device for pressure testing.

With water cooling device (can equipped with water cooler and external refrigeration equipment)

Control system

product-229-160

SIEMENS colors touch screen (6 inches), PLC control system, Bilingual in Chinese and English can be switched

Pressure sensor: -2500~2500Pa touch screen, accuracy±1Pa

Pressure control: control the box pressure to one setting point (scope can be set)

measurements

product-115-160product-115-160

Moisture analyzer: UK MICHELL brand, 0~1000ppm touch screen display, precision 0.1ppm

Oxygen analyzer: US Avcray brand, 0~1000ppm touch screen display, precision 0.1ppm

Vacuum Pump

product-229-229

Britain Brand EDWARDS, pumping speed 8.4-12m3/h, Model: RV12

Extreme pressure 2 x 10-3 mbar/1.5 x 10-3 Torr

Inlet flange NW25

Outlet flange NW25

Noise level: 50 Hz 时 48 dBA

Glove

product-229-160

Brand: US NORTH

Material: Butyl

Model for standard:8B1532

illumination system

Equipped with Fluorescent lamp. Brand: PHILIPS

Solvent removal system

Efficient adsorption of activated carbon, can effectively remove trace organic solvents;

Spare KF40 standard joint

Special design to KF40 Joint, you can lead the water into the box

product-229-160product-229-160

Can be connected to other devices (low temperature refrigerator,

film coating machine, microscope, solvent purification system, low temperature cold trap, heating furnace, etc.)

low temperature refrigerator

product-229-160product-229-160

microscope

product-229-160

Solvent Purification System

product-229-160

Coating machine

Inside connection(one door of Coating machine is inside of glove box)

product-229-160

Outside connection

product-229-160

Equipment size and room requirements

Overall size

3050mm(L) x 1000mm(W) x 1950mm(H)

120.08" x39.37" x 76.77"

Room size

2200mm(L) x 1500mm(W) x 2000mm(H)

133.86" x59" x 78.74"

Wooden Package size and weight

3385mm(L) x 1145mm(W) x 2300mm(H)

133.27" x 45" x 91" (5.8M3)

1020 kg (2197 lbs)

Notice

When the gas is circulating, must ensure that the piping of the circulation system is clear and cannot be blocked (if the color of the filter is changed from white to black, the filter should be replaced timely).

No corrosive gas in the glove box.

It needs to regenerate the column when the moisture and oxygen indicators cannot meet the work requirements.

If the corrosive gas is released from inside of glove box.(such as LiPF6 electrolyte), please do not use. you must put the bottle filled with liquid seal.

Vacuum pump needs to be checked the oil level and maintain in regular time.

 

2 work station

Battery Coating Machine

Battery Laboratory Equipment

Vacuum Mixer

Hot Tags: glove box with gas purification system, China glove box with gas purification system manufacturers, suppliers, factory